본문 바로가기
  • Home

A study for measurement of radial artery oxygen saturation system using photoelectric plenthysmography

  • Journal of The Korea Society of Computer and Information
  • Abbr : JKSCI
  • 2010, 15(3), pp.11-18
  • Publisher : The Korean Society Of Computer And Information
  • Research Area : Engineering > Computer Science

KIM KYUNGHO ORD ID 1

1단국대학교

Accredited

ABSTRACT

In this study, the pulse of radial artery and oxygen saturation are detected using photoelectric plethysmograph method. Using our device designed reflection type, we can detect the reflected light by radial artery and by switching circuit, we can also separate to 625nm band signal and 940nm band signals. The separated signals are converted as a pulse data by the pulse signal processing circuit. In this study, the reflection type of oxygen saturation calculation method is applied instead of the transmission type because of the reflection type sensor is used to measure the radial artery. As a result, we can detect about 97% accuracy of the oxygen saturation compare with the conventional method. For the accurate signals, the wrist band with sensor was designed and fixed on the radial artery. As a result, this wrist band type sensor was applicable to prevent position errors from motion artifact and could increase the accuracy during the measuring.

Citation status

* References for papers published after 2023 are currently being built.