본문 바로가기
  • Home

Comparison of the accuracy of digital models made from white light scanner by scanning method

  • Journal of Korean society of Dental Hygiene
  • Abbr : J Korean Soc Dent Hyg
  • 2012, 12(6), pp.1082-1089
  • Publisher : Korean Society of Dental Hygiene
  • Research Area : Medicine and Pharmacy > Dentistry

김기백 1 이경탁 1 Kim, Jae-Hong ORD ID 1

1고려대학교

Accredited

ABSTRACT

Objectives : The aim of this study was to determine the accuracy of digitized stone models, impression materials compared to the master model and the reliability of the computer aided analysis. Methods : A master model(500B-1, Nissin dental product, Japan) with the prepared lower full arch tooth was used. Ten vinyl polysiloxane impressions(Examix , GC Industrial Corp, Japan) of master model were taken and type Ⅳ stone(aesthetic-base gold , Dentona, Germany) were poured in stone models. The linear distance between the reference points were measured and analyzed on the Delcam Copycad (Delcam plc, UK). The t-student test for paired samples was used for statistical analysis. Results : The mean differences to master model for stone model and impression material were 0.11~0.19mm, and 0.19~0.29mm, respectively. There were statistical differences in dimensional accuracy for full arch impression between master model and stone model/impressions(p<.05). Conclusions : Two different scanning methods showed clinically acceptable accuracy of full arch digital impression produced by them. These results will have to be confirmed in further clinical studies.

Citation status

* References for papers published after 2023 are currently being built.